 TH 102 / TH 102-Ex
Rail mounted temperature transmitter, HART programmable, Pt 100 (RTD), thermocouples, electrical isolation
10/11-8.54 EN
I Input
- Resistance thermometer (2, 3, 4-wire circuit) - Thermocouples www..com - Resistance remote signalling unit (0...5000 ) - Voltages, mV (-125...1200 mV)
I Substitution strategy in case of error (NE43) I Approvals for explosion protection
- intrinsically safe - II 2 G EEx [ia] ib IIC T6, mount in zone 1 II 3 G EEx n A II T6, mount in zone 2
I Output
- 2-wire technique - 4...20 mA, HART signal
I I I I I
1 or 2 independent channals Electrcal isolation (I/O)) Digital low-drift processing of measurement values Customer-specific linearization Continuous sensor and self-monitoring - Parameter saved permanently in EEPROM - Monitoring of data integrety every 10 s
I Input functionality (absolute, differential, average value) I EMC acc. to EN 50082-2 and NE 21 I Parameterizationg
- PC software application SMART VISION - Hand held terminals STT 04, HC 275
I 5 years warranty
ABB Instrumentation
TH 102 / TH 102-Ex Rail mounted temperature transmitter, HART programmable
Technical data Output
Output signal (temperature linear) Residual ripple (peak-to-peak) Current consumption Max. output current Parameterizable current error signal Underranging Overranging Default value Damping 4...20 mA < 0.2 % < 3.6 mA 23.6 mA 3.6 mA 22 mA 3,6...22 mA t63 = 0...30 s
10/11-8.54 EN
Power supply
(2-wire methode: power supply wires=signal wires) Supply voltage, poling protected Us = 8.5...30 V DC for explosion protection application, max. Ui = 8.5...29.4 V DC Influence of supply voltage max. residual ripple < 0.05 %/10 V 1 % US (< 500 Hz)
Maximale Load
( Usmax - Usmin ) R ( k ) = ---------------------------------------------23, 6
Input
Resistance
Maximum permissible load R
910
Resistance thermometer (IEC 751, JIS, SAMA) min. span Max. line resistance (Rw) per core 2, 3, 4-wire Measuring current Sensor short-circuit Sensor break Input filter Thermocouples Types Voltages Min. span Sensor monitoring current Input filter Internal reference junction
www..com
n*Pt100/Ni100 to Pt1000/Ni1000; Cu (n=0.1; 0.2; 0.5; 1; 1.2; 2; 3...10) 15 K/50 K 7.5 , 10 , 50 300 A < 5 (for RTD) > 1.5 M 50/60 Hz
250
(HC)
O
t ra pe
io
n
n ra
ge
0 8,5
Z-20160_1
14,4
V DC Power supply Us
30
B, E, J, K, L, N, R, S, T, U -125 mV...125 mV -125 mV...1200 mV 2 mV / 50 mV 70 nA 50/60 Hz Pt 100, via software switchable (no jumper necessary)
Input element Sensor Thermocouple Type B Thermocouple Type E Thermocouple Type J Thermocouple Type K Thermocouple Type R Thermocouple Type S Thermocouple Type T Thermocouple Type N DIN 43710 Thermocouple Type L Thermocouple Type U IEC 751; JIS; SAMA1) Resistance thermometer Pt 100 2, 3 and 4-wire Resistance thermometer Pt 1000 DIN 437602) Resistance thermometer Ni 100 2, 3 and 4-wire Resistance thermometer Ni 500 Resistance Voltage mV Standard IEC 584-1
1) 2)
Measuring range 250...+1820 C (+482...+3308 F) -250...+1000 C (-418...+1832 F) -210...+1200 C (-346...+2192 F) 250...+1372 C (-418...+2502 F) - 50...+1768 C (- 58...+3215 F) - 50...+1768 C (- 58...+3215 F) -200...+1350 C (-328...+2462 F) -200...+1350 C (-328...+2462 F) -200... +900 C (-76...+ 482 F) -200... +600 C (-328...+1112 F) -200... +850 C (-328...+1562 F) -200... +850 C (-328...+1562 F) - 60... + 250 C (-76...+ 482 F) - 60... + 250 C (-76...+ 482 F) 0...500 / 0...5000 -125 mV...+125 mV -125 mV...+1200 mV
Min. measuring span 235 C (423 F) 30 C ( 54 F) 37 C ( 67 F) 54 C (98 F) 171 C (308 F) 193 C (348 F) 50 C ( 90 F) 60 C (108 F) 36 C ( 65 F) 40 C ( 72 F) 15 C ( 28 F) 50 C ( 90 F) 8 C ( 15 F) 15 C ( 28 F) 5 / 50 2 mV 50 mV
IEC 751 a = 0.00385; JIS a = 0,003916; SAMA a = 0,003902 Edison Curve No. 7
Page 2 of 6
03.2000
TH 102 / TH 102-Ex Rail mounted temperature transmitter, HART programmable
Technical data General characteristics
Response time Vibration resistance Vibration in operation Electrical isolation (I/O) Long-term stability < 0.5 s 2 g acc. to DIN IEC 68 part 2-6 1.5 kV AC 0.1 % p.a. Supply circuit Output [ib] Ui = 29.4 V Pi = 0.8 W Li = 220 H
10/11-8.54 EN
Explosion protection
Intrinsically safe Zone 1 EC certificate Temperature class T6/T5/T4 II 2 G EEx [ia] ib IIC T6 PTB99 ATEX 2139 X < 50 C/65 C/85 C Input [ia] Uo = 5.6 V Io = 145 mA3) Po = 20 mW Lo = 1 mH Co = 1.55 F II 3 G EEx n A II T6 PTB 99 ATEX 2216 X < 50 C/65 C/85 C
Environment conditions
Ambient temperature range (-40)-20...85 C Transport and storage temperature -40...100 C Relative humidity < 100 % (100 % humidity with isolated terminals only) condensation permitted
Max. voltage
Short-circuit current Ii = 130 mA Max. power Internal inductance
Internal capacitance Ci = 15 nF Zone 2
3) Load current for connected primary element [ia] < 1.5 mA
Mechanical construction
Dimensions Weight Housing material Type of protection Class of protection Class of combustibility Color (Epoxy) Electrical connection Terminals, pluggable 2.5 mm2, screw terminals cf. dimensional drawing 250 g Polyamid IP 20 (DIN 40050) 2 (IEC 348) V0 acc. to UL 94 light grey (RAL 9002)
Conformity declaration Temperature class T6/T5/T4
Electromagnetic compatibility (EMC)
Pt 100: measuring range 0...100 C, span 100 K Type of test burst to signal/ data lines static discharge contact discharge to: contact plate terminals for supply terminals for sensors radiated field 80 MHz...1 GHz coupling 150 kHz - 80 MHz Degree 2 kV Influence < 0.5 % IEC 1000-4-4 1000-4-2 8 kV 6 kV 3.75kV 10 V/m 10 V < 1.0 % < 1.0 % < 1.0 % 1000-4-3 < 1.0 % 1000-4-6 < 1.0 %
Characteristics at rated conditions*)
(acc. to IEC 770, related to 25 C) Measuring error incl. characteristic deviation Pt 100/resistance measurement < 0.2 % min. < 0.2 K/< 80 m Thermocouple/mV < 0.2 %, min. < 10 V
www..com junction internal reference
Additional influence of the
Acc. to NAMUR NE 21 recommendation Pt 100 DIN IEC 751 cl. B
Influences
Influence effect of temperature Pt 100/resistance measurement1)
ME () < (0.08 % + MS ()
x 0.008 %) / 10K
Thermocouple/mV2)
ME (mV) < (0.08 % + MS (mV)
x 0.01 % +
0.014 K MS (K)
x 100 %) / 10 K
Percentage related to measuring span MS = ME - MA MA = lower range value, ME = upper range value
*) Percentage related to set measuring span 1) Pt 100 (0...400 C): Effect of temperature influence < (0.08 % + 0.013 %)/10 K = 0.093 %/10 K
03.2000
Page 3 of 6
TH 102 / TH 102-Ex Rail mounted temperature transmitter, HART programmable
Communication/parameterization
Software-Tool SMART VISION Hand held terminals STT 04; HC 275 Parameter
10/11-8.54 EN
Sensor type, error signalling, measuring range, general characteristics (i. e. TAG number), damping, signal simulation of output Software interface AMS, Cornerstone
11 12 13 14 21 22 22 24
ABB
TH102-Ex
PTB 99 ATEX 2139 X II 2 G EEx [ia] ib IIC T6 TA<=50 C
Elektr. Daten siehe EG-Baumusterprufbesch.
Processinterface Line resistance
Ex
(250 min.)*
Kanal 1: PT 100-3L 0..100 C Kanal 2: PT 100-3L 0..100 C
41 42 31 32
or Isolators Controller Recorders ...
GND
or
(optional) FSK Modem
*if necessary *falls erforderlich
Z-20162e
1)
HC 275; STT04
1)
2-channel version: programming channel 2
www..com
Page 4 of 6
03.2000
TH 102 / TH 102-Ex Rail mounted temperature transmitter, HART programmable
Ordering information
TH 102 / TH 102-Ex Without Explosion protection: TH 102 1-channel TH 102 2-channel With Explosion protection TH 102-Ex 1-channel TH 102-Ex 2-channel Zone 1 ATEX: II 2 G EEx [ia]ib IIC T6 ATEX: II 2 G EEx [ia]ib IIC T6 Catalog No V115171 3 DM
10/11-8.54 EN
ex stock Lager
478,00 * 898,00 *
5 2
528,00 * 998,00 *
Zone 2 / type N TH 102-Ex N 1-channel ATEX: II 3 G EEx n A II T6 TH 102-Ex N 2-channel ATEX: II 3 G EEx n A II T6 Programming 1-channel Factory standard parameter Pt 100, 0...100 C, 4-wire circut, damping off, direct action characteristic overranging at sensor or device error ( 23,5 mA) Customer-specific parameter definition, two-channel version 2-channel Factory standard parameter Pt 100, 0...100 C, 4-wire circuit, damping off, direct action characteristic overranging at sensor or device error ( 23,5 mA) Customer-specific parameter definition, two-channel version Calibration certificates without 1-channel 2-point 1-channel 9-point 2-channel 2-point 2-channel 9-point
N M
528,00 * 998,00 *
0
1 2
40,00 *
3 0 1 2 A B
60,00 *
140,00 200,00 200,00 300,00 DM 618,00 480,00 530,00 50,00 50,00 200,00
Accessories
Catalog No H&B FSK modem [EEx ib] IIC (parameter setting in the installation) 0343705 Smart Vision5 Software on CD-ROM 7957777 31/2" Diskettes 7957778 Smart Vision5 manual German 7957779 English 7957780 TH 02 / -102 / -202 driver for AMS software 1.3.1 (Rosemount) 7957771 www..com 5 see Data Sheet 63-1.20 Minimum hardware reuqirements: Intel 80486; 66 MHz, 8MB RAM; free hard-disk capacity; Windows 3.x (95, 98, NT) * * * * * *
Notice: For a lokal programming on the desk can used as Hardware the universal TS02-Programming-Set (without Parasoft). (see Data Sheet 11-8.17 ordering information accessoriers )
03.2000
Page 5 of 6
TH 102 / TH 102-Ex Rail mounted temperature transmitter, HART programmable
Connection diagram
10/11-8.54 EN
Microprocessor 31 +
11 + Input
11
11
11
11
11
11
11
11
11
Us= 8,5...30V DC
Output Channel 1
12
-
K1
12 12 12 12
12 12
K1
12 12 12
13
13 13
13
13
K2
13
K2
13 13
4...20 mA
ReferenceJunction
32 14 21 + Input Channel 2
14 14 14 14 14
V DC V DC
14
V DC V DC
14
Microprocessor 41 +
21
21
21
21
21
21
Us= 8,5...30V DC
Output
22
-
K1
22 22 22
22
K1
22 22
23
23
23
23
K2
K2
23
4...20 mA
ReferenceJunction
42 24
24 24 24
V DC V DC
V DC V DC
24
a) For the 1 channel version view wiring diagram of channel 1
b)
c)
d)
e)
f)
g)
h)
i)
Z-20208
a) Resistance thermometer, 2-wire circuit b) Resistance thermometer, 3-wire circuit c) Resistance thermometer, 4-wire circuit d) Double resistance thermometer, 2-wire circuit e) Potentiometer input (4-wire circuit)
f) Thermocouple g) Double thermocouple h) Voltage measurement i) Double voltage measurement
Dimensional diagram (dimensions in mm)
11 12 13 14 21 22 22 24
Us= 8,5...30V DC
ABB
Microprocessor 31 + Output Channel 1 Input
11 +
11
11
11
11
11
11
11
11
11
www..com
99
TH102-Ex
PTB 99 ATEX 2139 X II 2 G EEx [ia] ib IIC T6 TA<=50 C
Elektr. Daten siehe EG-Baumusterprufbesch.
12
-
K1
12 12 12 12
12 12
K1
12 12 12
13
13 13
13
13
K2
13
K2
13 13
4...20 mA
ReferenceJunction
32 14 21 + Input Channel 2
14 14 14 14 14
V DC V DC
14
V DC V DC
14
Ex
Microprocessor 41 +
21
21
21
21
21
21
Us= 8,5...30V DC
Output
22
-
K1
22 22 22
22
K1
22 22
Kanal 1: PT 100-3L 0..100 C Kanal 2: PT 100-3L 0..100 C
23
23
23
23
K2
K2
23
4...20 mA
ReferenceJunction
42 24
24 24 24
V DC V DC
V DC V DC
24
a)
b)
c)
d)
e)
f)
g)
h)
i)
41 42 31 32
22,5
114,5
ABB Automation Products GmbH Sensycon Alzenau Borsigstrae 2 D-63755 Alzenau Tel. (0 60 23) 92 - 0 Fax (0 60 23) 92 - 33 00 http://www.abb.com/automation
Subject to technical changes. Printed in the Fed. Rep. of Germany Data Sheet 10/11-8.54 EN 03.2000
Z-20207
35


▲Up To Search▲   

 
Price & Availability of TH102

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X